UV SpectroPhotometer - Operation & Calibration

By: Pharma Tips | Views: 24849 | Date: 04-Sep-2011

A device which produces a spectrum of ultraviolet light and is provided with a calibrated scale for measurement of wavelength.To provide a procedure for operating the UV – VIS Spectrophotometer.

UV VISIBLE SPECTROPHOTOMETER
SOP FOR OPERATION

1.0. PURPOSE:
To provide a procedure for operating the UV – VIS Spectrophotometer.

2.0. SCOPE:
This covers the procedure for operating UV–VIS Spectrophotometer.
MODEL : UV-1700 with PC software
MAKE : Shimadzu.

3.0. RESPONSIBILITY:
3.1. QC-Chemist
3.2. QC-Officer

4.0. ACCOUNTABILITY:
Manager-Quality Control

5.0. DEFINITION:
A device which produces a spectrum of ultraviolet light and is provided with a calibrated scale for measurement of wavelength.

6.0. PROCEDURE:




6.1. For PHOTOMETRIC
6.1.1. Switch on the instrument and computer
6.1.2. Check the connection cable to the PC is properly connected or not.
6.1.3. Select the icon of UV probe and click double
6.1.4. Press PC control (F4) in the instrument.
6.1.5. Select the icon Photometry.
6.1.6. Click the CONNECT icon.
6.1.7. Place the blank solution in sample compartment
6.1.8. Click the method icon(M).The page will be opened enter the wavelength & add the parameter, select calibration and enter raw data and
          close the window and method should be saved in method folder( which is suitable for name of the sample).
6.1.9. Keep blank and perform AUTOZERO
6.1.10. To read the samples keep the test solution and enter the sample ID in sample table and click the Read unknown icon.(If the method
           is already exists open the method file place the blank and make AUTOZERO).
6.1.11. To take print out click on REPORT icon and go to file open , select “ B & B PHOTOMETRIC DISSOLUTION” .Change the necessary
           information and save the report in data folder as XYZ. Where X is name of the sample or code, Y is batch No & Z is date, month &
           data.
6.1.12. Then preview the report and give the print out command.
6.2. FOR SPECTRUM
6.2.1. Switch on the instrument and computer.
6.2.2. Check the connection cable to the PC is properly connected or not.
6.2.3. Select the icon of UV probe and click double.
6.2.4. Press PC control (F4) in the instrument.
6.2.5. Select the icon Spectrum and click.
6.2.6. Place the blank solution in sample compartment
6.2.7. Click the method icon(M).The page will be opened, enter the wavelength range & scan at medium speed, and enter the necessary
          information which are relevant then click and the method should be saved as name of the sample and do the base line correction. (If
          the method is already exists open the method file , place the blank and do the base line correction).
6.2.8. To read the sample spectrum place the sample solution in sample compartment and click the START icon.
6.2.9. After spectrum scanning completion new data set will appear on screen automatically. Enter the file Name :dzu\uvprobe\data\xyz
         where X is sample name or code, Y is B.No , Z is date and Month and data, data storage to be same enter the analyst name and
         enter the analyst name and comment if required, click ok and should be saved
6.2.10. After completion of scanning go to operation click on peak pick, pick table will appear on monitor.
6.2.11. To take print out click on REPORT icon and go to file open select “ B& B SPECTRUM PEAK PICK”
6.2.12. Change the necessary information allotted space and save the report
6.2.13. Then preview the report and then give print out command.

7.0. REFERENCES:

Instrument Manual

8.0. ENCLOSURES:

Nil

9.0. ABBREVIATIONS:

Nil


UV VISIBLW SPECTROPHOTOMETER

SOP FOR CALIBRATION

1.0.CONTROL OF ABSORBANCES

PROCEDURE:
REPARATION OF POTASSIUM DICROMATE SOLUTION: Dry K2Cr2O7 at 130oC for 2 hours. Weigh accurately about the quantity of sample equal to 60mg into 1000ml volumetric flask. Add 0.005m Sulphuric acid to dissolve and dilute with the same solvent.

Find out the absorbances of (A 1% 1cm) above said solution at wavelengths of a 235nm, 257nm, 313nm and 350nm. Calculate the (1% 1cm) value using the following expression.

Absorbance X 1 X 1000 / Weight of K2Cr2O7 in g X 100
Check the value for each wavelength with the tolerances given four wavelengths.


PREPARATION OF REAGENT:


(i) Dilute 3ml of concentrated sulphuric acid and make up 100ml with distilled water.
(ii) Dilute 20ml to 2000ml with distilled water.

2. LIMIT OF STRAYLIGHT:
The absorbance of a 1.2-% w/v solution of potassium chloride at a path-length of 1cm should be greater than 2.0 at about 200nm when compared with water as reference liquid.

1.2% POTASSIUM CHLORIDE SOLUTION:
Take 1.2 g of Potassium chloride previously dried at 105oC for two hours into the 100ml volumetric flask. Add water to dissolve and make up to volume with water.

3. RESOLUTION POWER:
Record the spectrum of a 0.02% v/v solution of Toluene in Hexane UV. The ratio of the absorbance at the maximum at about 269nm to that at the minimum at about 266nm is not less than 1.5 unless other wise specified in the monograph.
0.02% v/v

TOLUENE:
Take 0.02ml of Toluene into the 100ml volumetric flask. Add Hexane to volume.
Previous Page Next Page

People Searching On This Page:
  • standard operation procedure for spectrophotometer
  • spctrphotometer calibration sop
  • procedure for UV spectrophotometer
  • spectrophotometer operating procedure
  • procedures in operating a uv spectrometer
  • operating procedure of a spectrophotometer
  • uv calibration
  • uv visible spectrophotometer work instruction
  • what is operation and calibration of UV - visible spectrometer?
  • SOP for UV
  • operation of uv-vis spectrophotometer
  • nmt4
  • ultraviolet light safety procedures
  • uv calibration parameters
  • sop for calibration uv spectrophotometer

Related Pages


Muffle Furnace - Operation & Calibration

Muffle Furnace - Operation & Calibration

Articles | Pharmaceutical Equipment
Date:
09-Oct-2011  Views: 20586

To check the instrument performs satisfactorily and gives accurate temperature. ...
Vernier Calliper’s - Operation & Calibration

Vernier Calliper’s - Operation & Calibration

Articles | Pharmaceutical Equipment
Date:
09-Oct-2011  Views: 11530

To lay down procedure to operate and calibrate Digimatic Vernier Calliper. ...
Laminar Air Flow - Operation & Calibration

Laminar Air Flow - Operation & Calibration

Articles | Pharmaceutical Equipment
Date:
03-Oct-2011  Views: 35943

To lay down a Standard Operating Procedure for laminar flow. ...
Brookfield Viscometer - Operation & Calibration

Brookfield Viscometer - Operation & Calibration

Articles | Pharmaceutical Equipment
Date:
23-Sep-2011  Views: 25575

To provide a procedure for the operating Brookfield viscometer. ...
Centrifuge - Calibration and Operation

Centrifuge - Calibration and Operation

Articles | Pharmaceutical Equipment
Date:
30-Aug-2011  Views: 12643

To lay down an operating procedure and calibration of Centrifuge apparatus. ...
Post Your Comments (No Login Require)
Name : (required)
Email : (required)
Website :

Comment : (required)

31  + 7 =     
Comments
RJ  |  04-Sep-2011 20:07:46 IST
Why k band, r band so called to electronic transitions..
People Searched About:
Standard Operation Procedure For Spectrophotometer   |   Spctrphotometer Calibration Sop   |   Procedure For UV Spectrophotometer   |   Spectrophotometer Operating Procedure   |   Procedures In Operating A Uv Spectrometer   |   Operating Procedure Of A Spectrophotometer   |   Uv Calibration   |   Uv Visible Spectrophotometer Work Instruction   |   What Is Operation And Calibration Of UV - Visible Spectrometer?   |   SOP For UV   |   Ultraviolet Light Safety Procedures   |   Uv Calibration Parameters   |   UV Calibration Pharma   |   Spectrophotometer Operation   |   Cobicistat Uv Wave Length   |   Uv Calibration Procedure   |   Instrumental Start Up And Operation Spectrophotometer   |   Uv Spectrophotometer Operation   |   Operation Of Uv Vis Spectrophotometer   |   Standard Operating Procedures UV Shimadzu Spectrophotometer   |   Standard Operating Procedures For Shimadzu Spectrophotometer   |  
Google : 2604 times | Yahoo : 13 times | Bing : 1169 times |