Incubator Standard Operating Procedure

By: Pharma Tips | Views: 21658 | Date: 29-Aug-2011

To ensure that the incubator performs satisfactorily and maintain accurate temperature for Microbial growth.

SOP FOR OPERATION AND MAINTENANCE

1.0 OBJECTIVE
To ensure that the incubator performs satisfactorily and maintain accurate temperature for Microbial growth.

2.0 SCOPE

The SOP covers procedure for operation and calibration of incubator and is applicable to Quality Control Department formulation Units

3.0 RESPONSIBILITY

Microbiologist.

4.0 ACCOUNTABILITY

Department Head.

5.0 PROCEDURE


5.1 Ensure that the power supply to the incubator is switched ‘OFF’.
5.2 Dedust the incubator daily externally with a clean dry cloth.
5.3 Once a week remove adhered dust by wet mopping using detergent solution. Afterwards wipe the surface with a clean dry cloth to
      remove traces of detergent.
5.4 Once in a month clean the interior surfaces with 2.5 % savlon solution using a clean cloth.

6.0 OPERATING PROCEDURE
6.1 Ensure that the incubator is properly connected to the power supply.
6.2 Switch on the main .
6.3 Turn on the red colour power knob towards 0-1.
6.4 Turn on the cooling knob towards 0-1.
6.5 To set the incubator at 22°C , set the lower temperature 21 OC by pressing the ‘SET POINT -1’ and simultaneously adjust the
      temperature with the help of screw of SET and RST by screw driver.
6.6 .Set the higher temperature 23 OC by pressing the ‘SET POINT -2’ and simultaneously adjust the temperature with the help of screw of
       SET and RST by screw driver.
6.7 In the same manner the incubator can be set to 37°, 44° and 55°C whenever required by setting the lower temperature to 36°,43° and 54° C respectively and by setting the higher temperature to 38°,45° and 56° C respectively.
6.8 Record the temperature twice daily. i.e. in the morning and in the evening. The temperature should not differ ± 2° C from the set
     temperature.
7.0 Trouble shooting problems and remedial action :
7.1 The temperature display is not glowing. Check for power supply and proper electrical connections of instruments with power point.
7.2 Temperature is not even in the incubator. The air circulating fan may not be functioning. Check for it.
7.3 Report any discrepancy observed during operation or temperature monitoring to Q.C. Executive and notify the defect to Maintenance Department.

8.0 CALIBRATION

8.1 Set the incubator temperature to 22°C. Wait till the set temperature is reached.
8.2 Take a calibrated thermometer and dip it in a 500 ml beaker filled to 3/4 of the volume with Glycerol AR grade.
8.3 Keep the beaker inside, at the center of the incubator. close the incubator door. Allow the temperature to equilibrate for 30 minutes.
8.4 Observe the temperature shown by the thermometer. The temperature shown by the
display and by thermometer shall not differ by more than 0.5°C.
8.5 Record the temperature at two time intervals ( with a gap of 6 hours) in the temperature record.
8.6 By following the same procedure as above carry out calibration by setting the incubator temperature to 37°C, 44°C and 55°C.
8.7 Record any discrepancy observed during operation or during temperature monitoring to Quality Control Executive and notify the defect to technical assistant for rectification. Affix “BREAK DOWN” label on the instrument.

9.0 FREQUENCY OF CALIBRATION :

9.1 Once a month and after each maintenance job.
9.2 Record the calibration record in the format prescribed

Previous Page Next Page

People Searching On This Page:
  • operating principle of laboratory incubator
  • operating principle of incubator
  • principle of operation of an incubator
  • operating principle of the incubator for microorganism growth
  • operating priniciple of an incubator in microbiology
  • operating incubator
  • working procedure of incubator in laboratory
  • incubator operating procedure
  • autoclave sop
  • procedure to use incubator
  • nmt4
  • procedure to operate incubator laboratory
  • incubator slideshare principle and working
  • working incubator requirements
  • incubator for microbial growth

Related Pages


NMR Spectrometer : Standard Operating Procedure

NMR Spectrometer : Standard Operating Procedure

Articles | Analysis | Sprectroscopy
Date:
28-Nov-2013  Views: 2192

General Procedures For All Spectra,Loading the Sample,Locking the Instrument,Running a 1H Spectrum,Acquiring the Spectrum,Transforming the Data,Manipu ...
Fluid Bed Dryer Operating Procedure

Fluid Bed Dryer Operating Procedure

Articles | Pharmaceutical Equipment | Tablet
Date:
02-Oct-2011  Views: 10890

Fluid bed dryer also named fluidized bed dryer is one of our main products. To know its working principle, first you should know what "fluidizd bed" i ...
Lyophilizer - Standard Operating Procedure

Lyophilizer - Standard Operating Procedure

Articles | Pharmaceutical Equipment | Injection
Date:
02-Oct-2011  Views: 5029

The process of using a laboratory scale lyophilizer presents a number of unique hazards. These hazards include but are not limited to extreme pressure ...
Auto Clave Standard Operating Procedure

Auto Clave Standard Operating Procedure

Articles | Pharmaceutical Equipment
Date:
29-Aug-2011  Views: 5877

A chamber for sterilizing with steam under pressure. The original autoclave was essentially a pressure cooker. The steam tightened the lid. The device ...
Tablet Punch Press Operating Procedures

Tablet Punch Press Operating Procedures

Articles | Pharmaceutical Equipment | Tablet
Date:
27-Aug-2011  Views: 8012

Tablet Punch Press Safety and Operating Procedures, Tablet Punching Machine ...
Post Your Comments (No Login Require)
Name : (required)
Email : (required)
Website :

Comment : (required)

6  + 2 =     
Comments
ronak patel  |  07-Dec-2016 20:17:01 IST
nice blog..
yitbarek  |  14-Jun-2017 17:30:02 IST
sop is not clear specially about water pack
yitbarek  |  14-Jun-2017 17:32:49 IST
sop is not clear specially about water pack
FANEM
People Searched About:
Operating Principle Of Incubator   |   Operating Principle Of The Incubator For Microorganism Growth   |   Operating Priniciple Of An Incubator In Microbiology   |   Operating Incubator   |   Working Procedure Of Incubator In Laboratory   |   Autoclave Sop   |   Procedure To Use Incubator   |   Procedure To Operate Incubator Laboratory   |   Incubator For Microbial Growth   |   Incubator   |   Standard Operating Procedures For DNP-9022 Laboratory Incubator   |   Incubator Standard Operating Procedure   |   Incubator Calibration Procedure   |   Procedure For Calibration Of Incubator   |   Email Thedsteam.Com Loc:IN   |   Email Rst-Renault.Com.Ar Loc:IN   |   Incubator Monitoring Procedure   |   Sop For Incubator   |   Lyophilyzer Machine Maintaince Procedure   |   Operating Procedure For Incubator   |   Standard Operating Procedures AND Pharmaceutical   |  
Google : 2032 times | Yahoo : 13 times | Bing : 460 times |