Calibration For control of absorbances.Calibration For Limit of stray light

SOP FOR CALIBRATION

1.0 PROCEDURE : For control of absorbances.

1.1 Preparation of Potassium dichromate solution.

1.2 Dry Potassium dichromate at 130°C for 2 hours. Weigh accurately about 60mg of potassium dichromate into 1000ml volumetric flask. Add 0.005 M sulfuric acid to dissolve and dilute with same solvent.

1.3 Calibration :

Using the Wave lengths.

235 nm
313 nm
257 nm
350 nm

Find out the absorbance of a A(1% cm).
Calculate the A(1% cm) value as per the formula for each wavelengths.
Check the value for each wavelengths with tolerances given for all four wavelengths
Tolerances (ie)

Flame Photometer

2.0 PROCEDURE : For Limit of stray light

2.1 Preparation of 1.2% w/v of Potassium chloride solution.

2.2 Dry Potassium chloride at C for 2 hours. Weigh accurately about 1.2 g of potassium chloride into 100105 ml volumetric flask. Add water to dissolve and dilute with the same solvent.

2.3 Calibration :

Stray light detected at a wavelength at 200 nm

Find out the absorbance at 200 nm with tolerance should be greater than 2.0

3.0 Procedure for Resolution power

3.1 Preparation of 0.02% v/v solutions of toluene in hexane

3.2 Pipette 2 mL of Toluene AR grade to 100 ml with Hexane AR grade and mix.

3.3 Calibration :

Using the wavelengths
269 nm
266 nm
Find out the ratio of the absorbance at the maximum at about 269 nm to that at the minimum at about 266 nm with tolerance should be not less than 1.5.

4.0 Preparation of Reagents (0.005 M sulphuric acid)
i) Dilute 3 ml of concentrated Sulfuric acid to 100 mL with water and mix.
ii) To 10 mL solution into 1000 mL wirh water and mix.

5.0 DOCUMENTATION :
5.1 F:Sxxx/ UCR/A - UV-Visible Spectrophotometer Calibration record.

Previous Page Next Page

People Searching On This Page:
  • how to calibrate a flame photometer
  • procedures of calibrating flame photometers
  • calibration of flame photo meter
  • purpose of digital photometer
  • Callibration of the flame photometer
  • pharmaceuticul equipment mail
  • calibration and validation of flame photo metery
  • pharmasop for calibration of flame photometry
  • absorbance of plame photometer in pharmaceuticals
  • how to calibrate flame photometer
  • indirect calibration of flam photometer
  • calibration of flame photometre
  • S.O.P of digital flame photomeret
  • flame photometerbing 1 operation procedures
  • procedure preparation potassium dichromate

Related Pages


Moisture Balance - Operation & Calibration

Moisture Balance - Operation & Calibration

Articles | Pharmaceutical Equipment
Date:
14-Sep-2011  Views: 13706

To lay down a procedure for cleaning, calibration & Operation of the Analytical moisture balance. ...
Microscope Operation & Calibration

Microscope Operation & Calibration

Articles | Pharmaceutical Equipment
Date:
12-Sep-2011  Views: 8401

The purpose of this procedure is to describe the method of operating, cleaning and Calibration of Microscope. ...
Melting Point Apparatus - Operation & Calibration

Melting Point Apparatus - Operation & Calibration

Articles | Pharmaceutical Equipment
Date:
06-Sep-2011  Views: 13949

To lay down a procedure for operation and calibration of melting point apparatus. ...
Thermometer - Calibration

Thermometer - Calibration

Articles | Pharmaceutical Equipment
Date:
05-Sep-2011  Views: 5512

To provide a procedure for calibration of thermometers with standard thermometer. ...
UV SpectroPhotometer - Operation & Calibration

UV SpectroPhotometer - Operation & Calibration

Articles | Pharmaceutical Equipment
Date:
04-Sep-2011  Views: 22308

A device which produces a spectrum of ultraviolet light and is provided with a calibrated scale for measurement of wavelength.To provide a procedure f ...
Post Your Comments (No Login Require)
Name : (required)
Email : (required)
Website :

Comment : (required)

24  + 5 =     
Comments
People Searched About:
How To Calibrate A Flame Photometer   |   Procedures Of Calibrating Flame Photometers   |   Calibration Of Flame Photo Meter   |   Purpose Of Digital Photometer   |   Callibration Of The Flame Photometer   |   Pharmaceuticul Equipment Mail   |   Calibration And Validation Of Flame Photo Metery   |   Pharmasop For Calibration Of Flame Photometry   |   Absorbance Of Plame Photometer In Pharmaceuticals   |   How To Calibrate Flame Photometer   |   Flame Photometerbing 1 Operation Procedures   |   Procedure Preparation Potassium Dichromate   |   Sulphuric Acid Calibration Using Uv   |   Standerd Preperation In Flame Photometer   |   TEXT   |   Calibration Of Flamephotometry   |   Flame Photometer Procedure   |   Calibration Of Flame Photometer   |   Flame Sxxx   |   Potassium Photometer Procedure   |   Potassium Dichromate Solution For Calibration   |  
Google : 1222 times | Yahoo : 12 times | Bing : 508 times |