BOD Incubator - Calibration

By: Pharma Tips | Views: 7492 | Date: 29-Aug-2011

This SOP provides an authorized procedure for cleaning, calibration and validation of BOD incubator.

SOP FOR CALIBRATION

1.0 .PURPOSE:
This SOP provides an authorized procedure for cleaning, calibration and validation of BOD incubator.

2.0 SCOPE :

This SOP covers the cleaning, calibration and validation method for BOD incubator (QCI – 43).

3.0 RESPONSIBILITY :

Microbiologist.

4.0 PROCEDURE

4.1.0 CALIBRATION :
Calibration of temperature indicator view controller to be carried out once in a year by external agency or whenever any maintenance done on the apparatus.
4.1.1 :Maintain the record of calibration.

4.2.0 VALIDATION
4.2.1 :Perform the validation of B.O.D. Incubator once in a year.
4.2.2 :Follow the operating procedure for operation after altering set temperature perform the temperature mapping inside the chamber.
4.2.3 :Five calibrated thermometers are kept in different places inside the incubator.
4.2.4 :All the thermometers should maintain the temperature of
± 2 °C of set temperature for 8 hours.
4.2.5 :Maintain the records of validation.

4.3.0 CLEANING METHOD
4.3.1 Disconnect the BOD Incubator from main electrical supply.
4.3.2 Clean the outside and inside of incubator with dry non shredding cloth. Wipe with IPA 70% v/v.
4.3.3 Maintain the log book for cleaning.
4.3.4 Frequency : Once in a week

Previous Page Next Page

People Searching On This Page:
  • pharmaceuticul equipment mail
  • BOD INCUBATOR SOP
  • STANDARD OPERATING PROCEDURE FOR AN INCUBATOR
  • working efficiency and calliberation of BOD incubator
  • validation of bod incubator
  • liquid used to clean BOD incubators
  • sop for bod incubator
  • Validation and Calibration of incubator
  • bod incubator calibration method in Bp
  • SOP clean for the gLycerine
  • Sop BOD
  • validation of sop of incubator
  • calibration of bod incubator
  • use of glycerine to calibrate incubator
  • CLEANING PHARMACEUTICAL EQUIPMENT WITH IPA

Related Pages


Digital Balance - Operation & Calibration

Digital Balance - Operation & Calibration

Articles | Pharmaceutical Equipment
Date:
03-Sep-2011  Views: 23419

Standard Operating Procedure:To ensure correct method of operation for balance. ...
pH Meter - Operation & Calibration

pH Meter - Operation & Calibration

Articles | Pharmaceutical Equipment
Date:
03-Sep-2011  Views: 10389

To lay down a procedure for operation of pH¬ meter (Model pH scan-2). ...
Disintegration Tester - Operation & Calibration

Disintegration Tester - Operation & Calibration

Articles | Pharmaceutical Equipment | Tablet
Date:
01-Sep-2011  Views: 21316

Tablet disintegration testing is used as a quality-assurance measure. It is not a true predicter of how well the dosage form will release its active i ...
Centrifuge - Calibration and Operation

Centrifuge - Calibration and Operation

Articles | Pharmaceutical Equipment
Date:
30-Aug-2011  Views: 9752

To lay down an operating procedure and calibration of Centrifuge apparatus. ...
BOD Incubator - Operation

BOD Incubator - Operation

Articles | Pharmaceutical Equipment
Date:
29-Aug-2011  Views: 6093

This SOP provides an authorized procedure for operation ...
Post Your Comments (No Login Require)
Name : (required)
Email : (required)
Website :

Comment : (required)

13  + 3 =     
Comments
Arpit  |  28-Dec-2016 15:12:14 IST
Thanks for the valuable instructions regarding bod incubator and get good quality bod incubator suppliers for laboratories and research works.
People Searched About:
Pharmaceuticul Equipment Mail   |   BOD INCUBATOR SOP   |   STANDARD OPERATING PROCEDURE FOR AN INCUBATOR   |   Working Efficiency And Calliberation Of BOD Incubator   |   Validation Of Bod Incubator   |   Liquid Used To Clean BOD Incubators   |   Sop For Bod Incubator   |   Validation And Calibration Of Incubator   |   Bod Incubator Calibration Method In Bp   |   Use Of Glycerine To Calibrate Incubator   |   CLEANING PHARMACEUTICAL EQUIPMENT WITH IPA   |   Oxygen Incubator Calibrators   |   Steps Of Calibration For Incubator   |   Email Ipa.Br Loc:IN   |   Incubator Calibration Procedure   |   How To Calibrate Incubator   |   How To Calibrate A BOD Incubator   |   SOP For Incubator Cleaning   |   Calibration Of A Incubator   |   BOD INCUBATOR CALIBRATION FORM   |  
Google : 888 times | Yahoo : 9 times | Bing : 474 times |