To lay down a procedure for cleaning, calibration & operation of the analytical balance (METTLER AR 104).

SOP FOR CLEANING, CALIBRATION AND OPERATION

1.0 OBJECTIVE

To lay down a procedure for cleaning, calibration & operation of the analytical balance (METTLER AR 104).

2.0 SCOPE
Module-4.

3.0 RESPONSIBILITY
Production officer.
QA chemist.

4.0 ACCOUNTABILITY
Production Executive.

5.0 PROCEDURE
5.1. Cleaning.
5.2. Calibration.
5.3. Operation.
5.4 Shut down.



5.1 Cleaning

5.1.1 Ensure that the power supply to the balance is switched ‘off’ before cleaning.
5.1.2 Clean the pan and inside of the balance with brush.
5.4.1 Clean the outside of the balance with clean and dry lint free duster every day.

5.2 Calibration

5.2.1 Frequency of calibration
5.2.2 Self Calibration Procedure

Frequency daily and after each maintenance job.
5.2.2.1 Clean the balance as per procedure given in 5.1
5.2.2.2 Check that air bubble is at the center of the level indicator and if required adjust the level by turning the leveling screws.
5.2.2.3 Switch on the balance and allow it to stabilize for 30 minutes.
5.2.2.4 Press the ‘cal’ button.The display should read ‘CAL’ followed by the blinking of 100.0000
5.2.2.5 Place 100.00g standard weight (METTLER).The display will blink 0.0000 g. Remove standard weight from balance.
5.2.2.6 After a few seconds the display should read ‘CAL DONE. This denotes end of self calibration.

5.2.3 Calibration against Standard weights

Frequency every month and after each maintenance job.
5.2.3.1 Clean the balance as per procedure given in 5.1
5.2.3.2 Check the weight of the standard weights 0.1000g, 0.2000g, 0.5000 g & 1.000 gms
5.2.3.3 Record the observations in calibration record.
5.2.3.4 If the variation is more than the least count label the balance ‘out of order’.
5.2.3.5 Error must be rectified and balance should be calibrated before use.

5.3. Operation

5.3.1 Ensure that balance is properly connected to the power supply.
5.3.2 Check that the air bubble of level indicator is at the center, and if required adjust the level by turning the leveling screws.
5.3.3 Switch on the main switch, this balance displays ‘88888888’ followed by 0.0000 g.
5.3.4 Place a clean petridish, close the balance door and press tare button on the bar to tare the petridish weight. The display should read
        0.0000g.
5.4.1 Place the required No. of tablets/capsules on petridish and close the balance door.
5.3.6 Wait until the stability symbol “O” displayed, then read of weight.

5.3 Shout down

5.4.1 Clean the balance as per procedure given in 5.1
5.4.2 Switch ‘off’ the balance.
5.4.3 Switch ‘off’ the mains

Previous Page Next Page

People Searching On This Page:
  • mettler balance calibration procedure
  • sop for analytical balance
  • sop of analytical balance
  • SOP verification of Balance
  • good Calibration sop of Analytical Balance
  • ICH analytical balances
  • mettler balance cleaning
  • pharma balance daily check
  • calibration of balance weights fda ich
  • analytical equipments
  • balance verification pharamceutical
  • calibration of balance procedure
  • ICH, analytical balance
  • Calibration of analytical balance SOP
  • inside analytical balances

Related Pages


QA Bioanalytical

QA Bioanalytical

Jobs | Quality Assurance
Date:
12-Jul-2011  Views: 6546

Team Lead-Quality Assurance Requirement for Bioanalytical:B Pharm/ M. Pharm along with minimum 4 year experience in QA-Bioanalytical.Knowledge and und ...
Chemist - B.sc / M.sc - Shraddha Analytical Services

Chemist - B.sc / M.sc - Shraddha Analytical Services

Jobs | Quality Control | Mumbai
Date:
27-Jun-2011  Views: 9833

SHOULD BE WELL VERSED WITH THEORITICAL AND PRATICAL KNOWLEDGE IN INSTRUMENTS SUCH AS GC, HPLC AND AAS. SHOULD KNOW TRITRIMETRIC ANALYSIS ...
Trainee - Analytical Chemist - MERCK LTD

Trainee - Analytical Chemist - MERCK LTD

Jobs | Quality Control | Goa
Date:
23-Jun-2011  Views: 10361

• In 1967, Merck was set up as the first Merck subsidiary company in Asia, currently it has a manufacturing facility at Goa. Merck Limited went on to ...
Vice President R&D (Formulations & Analytical Development API )

Vice President R&D (Formulations & Analytical Development API )

Jobs | R & D | Mumbai
Date:
01-May-2011  Views: 5826

Project Planning, Feasibility, execution at various levels and evaluation. Translate developmental plans to operational activities. Provide broad expe ...
Sr.General Manager - Analytical Research (APIs)

Sr.General Manager - Analytical Research (APIs)

Jobs | R & D | Ahmedabad
Date:
26-Mar-2011  Views: 6405

Head API Analytical Research with large team reporting to Head R&D (President) ...
Post Your Comments (No Login Require)
Name : (required)
Email : (required)
Website :

Comment : (required)

21  + 4 =     
Comments
People Searched About:
Mettler Balance Calibration Procedure   |   Sop For Analytical Balance   |   Sop Of Analytical Balance   |   SOP Verification Of Balance   |   Good Calibration Sop Of Analytical Balance   |   ICH Analytical Balances   |   Mettler Balance Cleaning   |   Pharma Balance Daily Check   |   Calibration Of Balance Weights Fda Ich   |   Analytical Equipments   |   Balance Verification Pharamceutical   |   Calibration Of Balance Procedure   |   ICH, Analytical Balance   |   Calibration Of Analytical Balance SOP   |   Inside Analytical Balances   |   Analytical Balance-Mettler   |   Balance Procedure Analytical Pharmaceutical   |   Sop For Balance Calibration   |   Analytical Balance By Merck   |   Self Calibrating Balance Mettler   |   Cleaning Analytical Weights   |  
Google : 674 times | Yahoo : 13 times | Bing : 754 times |